BrainBit for developers Subscribe for updates Visit website

DragonEEG

PURPOSE OF THE DRAGONEEG

The DragonEEG system is designed for recording brain biopotentials, with the capability of simultaneously recording signals from polygraphic channels. The system records electroencephalograms (EEG) with up to 21 channels, according to the generally accepted "10-20%" or "10-10%" EEG electrode placement schemes.

However, it does not provide report generation for diagnostic purposes. The system is not intended for clinical research.

The DragonEEG system is intended for use in:

  • Psychologists' offices
  • Research laboratories
  • Universities
  • Rehabilitation centers
  • Other organizations involved in brain research and neurotechnology development

The DragonEEG Wireless Brain Biopotential Monitoring System is safe and reliable. It complies with international and national standards and meets the requirements necessary for its intended use as defined by the manufacturer.

You can try our Demo App (Android), Demo App (iOS).

 

TECHNICAL CHARACTERISTICS

Characteristic

Value

Connectivity with mobile device or PC

Bluetooth LE 5.0

Number of recorded channels

24

Sampling frequency

500 Hz

ADC resolution

24 bits

Amplifier input impedance

1000 MΩ

Common-mode rejection ratio

110 dB at 50 Hz

DC current in the client circuit during signal measurement for each electrode

no more than ±1nA

Noise level on shorted inputs

no more than 2 μV peak-to-peak

DC current in the client circuit during inter-electrode resistance assessment

no more than ±10 μA

Battery charge time

up to 4 hours

Duration of device operation on battery

up to 8 hours

The battery charger with USB cable plugs into any power source port compatible with USB version 1.1 or later.

 

GENERAL INFORMATION ABOUT THE DEVICE

List of components basic system delivery set with dry electrodes

No.

Item

Quantity

1.

Amplifier DragonEEG

1

2.

Cable system I

1

3.

Cable system II

1

4.

Ear clip for referential electrode

2

5.

SD card

1

6.

SD card reader

1

7.

USB charging cable

1

8.

Cap size S

1

9.

Cap size M

1

10.

Cap size L

1

11.

Сase

1

12.

Bluetooth adapter

1

List of components basic system delivery set with wet electrodes

No.

Item

Quantity

1.

Amplifier DragonEEG

1

2.

Cable system I

1

3.

Cable system II

1

4.

Ear clip for referential electrode

2

5.

SD card

1

6.

SD card reader

1

7.

USB charging cable

1

8.

Cap size S

1

9.

Cap size M

1

10.

Cap size L

1

11.

Сase

1

12.

Bluetooth adapter

1

 

DragonEEG Operating Modes and LED Indicator

The DragonEEG device operates in three distinct modes: Turned off, Turned on, and Pairing mode.

1. Turned off: The device is completely powered down, and no LED indicators are active.
2. Turned on: To turn on the DragonEEG, simply click the power button. Once activated, the power button should be immediately surrounded by a solid red light. In this mode, the device is operational but not actively seeking a new Bluetooth connection.
3. Pairing mode: To put your DragonEEG into pairing mode, first ensure the device is turned on. Then, hold down the power button for approximately 5 seconds. When the device successfully enters this mode, the LED indicator on the DragonEEG should begin to flash quickly. This rapid flashing is the visual confirmation that the device is ready for a connection attempt.

To turn off the DragonEEG device, simply click the power button while it is in either the Turned on or Pairing mode.

Device indication and operating modes:

  • LED off: The recording module is off.
  • LED flashes 3 times: The recording module is powering on.
  • LED is on: The recording module is on and in standby mode.
  • LED flashes continuously (0.1s on, 0.7s off): Signal recording is in progress.
  • LED flashes quickly: Device is in pairing mode and ready for a connection attempt
  • LED flashes (0.25s on, 0.25s off): The recording module is charging.
  • LED is on (when the power supply is connected): The recording module is fully charged.

To turn the recording module on/off, press and hold the button for 2 seconds.